Facility: | Chair of Materials Science and Nanotechnology | Partner: | TU Dresden |
Location: | HAL 004 |
Microvolume spectrophotometer
Absorbance Accuracy 3% (at 0.74 Abs at 350nm)
Certifications/Compliance UL/CSA and CE
Detector Type 2048-element linear silicon CCD array
Wavelength Accuracy ±1nm
Absorbance Range Pedestal: 0-300 ABS Cuvette: 0 - 1.5 ABS
Accuracy 0.002
Accuracy (Absorption) 3% at (0.74 at 350nm)
Applications Nucleic Acid Quantification, DNA Quantification, RNA Quantification, Protein Quantification
Compatibility Microsoft Windows 7 Professional (32-bit and 64-bit), Windows 8 (32- and 64-bit), and Windows 10 Pro (64-bit)
Concentration 2ng/µl-15,000ng/µl (dsDNA) ng/µl
Connections USB
Depth (Metric) 20cm
Description Microvolume Spectrophotometer, with cuvet capability
Detection Limits Pedestal: 2ng/µL (dsDNA), 0.10mg/mL (BSA) Cuvette: 0.4ng/µL (dsDNA), 0.01 mg/mL (BSA)
Detection Range Pedestal: 2-15,000ng/µL (dsDNA) , 0.10 - 400mg/mL (BSA)
Footprint 14 x 20cm
Heating Range 37°; ±0.5°C
Includes n.a.
Item Description NanoDrop 2000c
Lamp Xenon Flash
Measurement Time <3 sec.
No. of Samples 1
Pathlength (Metric) 10; 5; 2; 1mm
Power Consumption 12VDC
Sample Volume (Metric) 1000, 0.5-2.0µL
Spectral Resolution =1.8nm (FWHM at Hg 253.7)
System Requirements Microsoft Windows 7 Professional (32-bit and 64-bit), Windows 8 (32- and 64-bit), and Windows 10 Pro (64-bit)
Type Spectrophotometer
Volume (Metric) 0.5 to 2.0µL
Voltage 12VDC
Wattage 5W
Wavelength Range 190-840nm
Weight (English) 2.1kg
Weight (Metric) 2.0kg
Width (English) 14cm
Width (Metric) 14cm
Warranty One year parts and labor
Unit Size Each
Facility: | Chair of Materials Science and Nanotechnology | Partner: | TU Dresden |
Location: | HAL 004 |
Microvolume spectrophotometer
Absorbance Accuracy 3% (at 0.74 Abs at 350nm)
Certifications/Compliance UL/CSA and CE
Detector Type 2048-element linear silicon CCD array
Wavelength Accuracy ±1nm
Absorbance Range Pedestal: 0-300 ABS Cuvette: 0 - 1.5 ABS
Accuracy 0.002
Accuracy (Absorption) 3% at (0.74 at 350nm)
Applications Nucleic Acid Quantification, DNA Quantification, RNA Quantification, Protein Quantification
Compatibility Microsoft Windows 7 Professional (32-bit and 64-bit), Windows 8 (32- and 64-bit), and Windows 10 Pro (64-bit)
Concentration 2ng/µl-15,000ng/µl (dsDNA) ng/µl
Connections USB
Depth (Metric) 20cm
Description Microvolume Spectrophotometer, with cuvet capability
Detection Limits Pedestal: 2ng/µL (dsDNA), 0.10mg/mL (BSA) Cuvette: 0.4ng/µL (dsDNA), 0.01 mg/mL (BSA)
Detection Range Pedestal: 2-15,000ng/µL (dsDNA) , 0.10 - 400mg/mL (BSA)
Footprint 14 x 20cm
Heating Range 37°; ±0.5°C
Includes n.a.
Item Description NanoDrop 2000c
Lamp Xenon Flash
Measurement Time <3 sec.
No. of Samples 1
Pathlength (Metric) 10; 5; 2; 1mm
Power Consumption 12VDC
Sample Volume (Metric) 1000, 0.5-2.0µL
Spectral Resolution =1.8nm (FWHM at Hg 253.7)
System Requirements Microsoft Windows 7 Professional (32-bit and 64-bit), Windows 8 (32- and 64-bit), and Windows 10 Pro (64-bit)
Type Spectrophotometer
Volume (Metric) 0.5 to 2.0µL
Voltage 12VDC
Wattage 5W
Wavelength Range 190-840nm
Weight (English) 2.1kg
Weight (Metric) 2.0kg
Width (English) 14cm
Width (Metric) 14cm
Warranty One year parts and labor
Unit Size Each